This regulation applies to the initial verification, continuous verification and in-use inspection of carbon monoxide detection alarms (hereinafter referred to as instruments) for detecting carbon monoxide gas concentrations in non-mine working environments. Instruments can be divided into instruments for continuous measurement and instruments for discontinuous measurement. JJG 915-2008 Verification Regulation for Carbon Monoxide Detection Alarms JJG915-2008 Standard download decompression password: www.bzxz.net
This regulation applies to the initial verification, continuous verification and in-use inspection of carbon monoxide detection alarms (hereinafter referred to as instruments) for detecting carbon monoxide gas concentrations in non-mine working environments.
Instruments can be divided into instruments for continuous measurement and instruments for discontinuous measurement.
1 Scope (1) 2 Overview (1) 3 Metrological performance requirements (1) 3.1 Measuring range (1) 3.2 Indication error (1) 3.3 Repeatability (1) 3.4 Response time (1) 4 General technical requirements (1) 4.1 Appearance (1) 4.2 Inspection of alarm function and alarm setting value (2) 5 Control of measuring instruments (2) 5.1 Verification conditions (2) 5.2 Verification items (2) 5.3 Verification method (2) 5.4 5.5 Calibration cycle (4)
Appendix A Carbon monoxide detection alarm calibration record (5)
Appendix B Inner page of calibration certificate (6)
Appendix C Inner page of calibration result notice (7)
Some standard content:
National Metrology Verification Regulation of the People's Republic of China JIG.915 Carbon Monoxide Detectors 2008 12 ~ 22 Issued 2009 06 - 22 Implementation General Administration of Quality Supervision, Inspection and Quarantine·Issued JJG915—2008 Verification Regulation of Carbon Monoxide Detectors JJG 915---2008 Replaces JJG915--1996 This regulation was approved by the General Administration of Quality Supervision, Inspection and Quarantine on December 22, 2008, and took effect on June 22, 2009. Responsible unit: National Technical Committee for Metrology of Environmental Chemistry Drafting unit: Shanghai Institute of Metrology and Testing Technology Participating drafting unit: Baoshan Iron & Steel Co., Ltd. This regulation entrusts the National Technical Committee for Metrology of Environmental Chemistry to be responsible for the interpretation of this regulation. Main drafters; 3JG 5-2008 Cai Jianhua (Shanghai Institute of Metrology and Testing Technology) Participating drafters: Wang Xiaoyan (Shanghai Institute of Metrology and Testing Technology) Chen Lan (Shanghai Institute of Metrology and Testing Technology) Zhu Xinren (Baoshan Iron & Steel Co., Ltd.) 1 2Overview· 3"Metrological performance requirements Measurement application 3.2Indication error 3.3Repeatability 3.4Response time 3.5Drift 4General technical requirements 4.1Appearance· JJG 915—2008 4.2 Inspection of alarm function and alarm setting value 5 Control of measuring instruments- Verification conditions 5.2 Verification items…. 5.3 Verification methods…. 5.4 Processing of verification results Verification cycle- Appendix B Appendix C -Verification record of carbon dioxide detection alarm Inside page of verification certificate Inside page of verification result notice (1) (1) 1 Scope JIG 915—208 Carbon Monoxide Detection Alarm Calibration Procedure This procedure is applicable to the initial calibration, subsequent calibration and in-use inspection of carbon monoxide detection alarms (hereinafter referred to as instruments) for detecting carbon monoxide gas concentrations in non-mining working environments. Instruments can be divided into continuous measurement instruments and non-continuous measurement instruments. 2 Overview The instrument is mainly composed of a sensor plus electronic components and a display part. The sensor converts carbon monoxide gas in the environment into an electrical signal, which is then processed by electronic components and displayed as a concentration value. According to the sampling method, the instrument can be divided into quasi-dispersion type and suction type. 3 Metrology performance requirements 3.1 Measurement range (0~2 000)μmol/mol. 3.2 Indication error Absolute error: ±5μmol/mol; Relative error: ±10%, It is sufficient to meet the above. 3.3 Repeatability 3.4 Response time Diffusion type ≥60s: Absorption type ≤30%. 3.5 Drift 3.5.1 Zero drift: 3umol/mol. 3.5.2 Range drift: ±5%. General technical requirements 4.1 Appearance 4.1.1 The instrument shall be marked with the name of the manufacturer, instrument model and number, manufacturing date, manufacturing license mark and number of the measuring instrument, and the accessories shall be complete: and the instruction manual shall be attached. 4.1.2 The display of the instrument shall be clear and complete. Each regulator component should be able to work normally, and each fastener should not be loose. 4.1.3 The instrument should not have any external damage that affects its normal operation. The coating of the newly manufactured instrument should not have obvious color scratches and peeling, and the joints of each component should be flat. 4.1.4. Diffusion instruments should be equipped with special label covers. 1 4.2 Inspection of alarm functions and alarm setting values JJG915—2008 4.2.1 After the instrument is turned on, the sound or light alarm should display normally. 4.2.2 Check the alarm setting value of the instrument. 4.2.3 If the instrument is set with more than two alarm setting values, mainly check the lower limit alarm setting value of the instrument. 5 Measuring instrument control The measuring instrument control of the instrument includes initial verification, subsequent verification and in-use inspection. ENE 5.1 Verification conditions : 111Ambient temperature: (0GS7OH 5.1.1 Verification environment conditions 5.1.1.2 Relative humidity: ≤8# 5.1.1.3 There should be no electromagnetic interference that affects the normal operation of the instrument. 5.1.2 Verification equipment 5.1.2.1 Gas standard material Cyclone gas standard material ( monoxide in air hereinafter referred to as standard gas), its expanded uncertainty should not be greater than 2.0% (k=2). 5.1.2.2 Zero point gas High purity nitrogen (purity not less than 99.999% 5.1.2.3 Flow rate Controller The flow controller consists of two gas flow meters, as shown in Figure 1. The gas flow meter range is: (0~1000)mL/min, and the accuracy level is not less than level 4. 5.1.2.4 Stopwatch: Resolution ≤0.1S. 5.2 Verification items Verification items are shown in the table Verification items Alarm concentration value Alarm function Indication error Repeatability Response time First verification List of verification items Continuous verification ETROLOG Note: 1 "+" means items to be inspected, "-" means items that can be inspected Items not to be inspected 2 After the instrument has been repaired and its main components have been replaced, it should be inspected in accordance with the requirements for the initial inspection 5.3 Inspection method 5.3.1 Appearance Use visual inspection and hand feel method in accordance with the requirements of 4.1. 2 Inspection during use 5.3.2 Adjustment of the instrument JJG915--2008 Preheat and stabilize the instrument and adjust the zero point and indication according to the requirements of the instrument manual When calibrating the instrument, connect the standard gas, flow controller and the instrument to be inspected as shown in Figure 1. Use the flow controller to control the flow of the standard gas according to the sampling method of the instrument to be inspected. When calibrating a diffusion instrument, the flow rate should be based on the requirements of the instrument manual. If the instrument manual does not clearly require If the requirement is met, it is generally controlled within the range of (200 ± 50) mL/min: When calibrating the suction instrument, it must be ensured that the bypass flowmeter in the flow controller has a flow discharge flow controller flow controller inlet SIFONMSL tested instrument standard gas 5.3.3 Alarm setting value and alarm function inspection instrument startup stability, pass a concentration of about instrument alarm lower limit 5.3.4 Indication value Pass the flowmeter 1.5 times the instrument alarm set value, record (lower limit) set value and observe whether the instrument sound or light alarm is normal. For the instrument , pass the first calibration and subsequent calibration with a concentration of about in turn. Use zero gas to adjust the zero point of the instrument, and use standard gas with 1.5 times the instrument alarm lower limit) setting value, 30% of the upper limit of the measurement range, and 70% of the upper limit of the measurement range. Record the actual reading of the instrument after the gas is passed through and repeat the measurement. Record the instrument display value A respectively. According to formula and 4, take the 4 with the largest absolute value. Sum (1) and formula (2 half calculate the indication difference of the instrument at the three concentration test points as the indication error of the instrument. Where: 4. Relative error A—Absolute error AA.X100% A—The arithmetic mean of the instrument reading values at the three concentration test points: LOH A,·Concentration value of standard gas For the inspection of the instrument during use, the instrument's set value should be determined first, and the concentration should be selected to be approximately 1.1 times the standard gas is good and the zero point and indication are adjusted. After the standard gas is passed in, record the display value of the instrument, measure 3 times, and calculate the indication error of the instrument according to formula (1) or formula (2). 5.3.5 Repeatability Use zero point gas to adjust the zero point of the instrument, pass in standard gas with a concentration of about 70% of the upper limit of the measurement range, and after the reading is stable, record the instrument display value A. Repeat the above measurement 6 times, and the repeatability is expressed as the relative standard deviation of a single measurement. Calculate the repeatability S of the instrument according to formula (3): 1 >(A, -A)2 Where: A,--instrument reading value; JJG 915--2008 A-the arithmetic mean of the instrument reading value; n--the number of measurements (n=6). 5.3.6 Response time For the first calibration and subsequent calibration of the instrument. Use zero gas to adjust the zero point of the instrument, pass standard gas with a concentration of about 70% of the upper limit of the measuring range, remove the standard gas after reading the stable value, pass zero gas until the instrument stabilizes, then pass standard gas with the above concentration, and use a stopwatch to record the time from the moment the standard gas is passed to the time when the instrument displays 90% of the stable value. Repeat the measurement three times, and take the average of the three measurements as the response time of the instrument. For the in-use inspection of the instrument. While checking the indication error of the instrument in use, calibrate the response time of the instrument. Measure twice and take the average value as the response time of the instrument. 5.3.7 Drift After passing zero gas until the instrument stabilizes, record the instrument display value A, then pass standard gas with a concentration of about 70% of the upper limit of the measuring range. After the instrument stabilizes, record the reading Am, and remove the standard gas. The non-continuous measurement instrument runs continuously for 1 hour: repeat the above steps every 15 minutes. The continuous measurement instrument runs continuously for 4 hours, repeat the above steps every 1 hour, and record the instrument display values A and A (i=1, 2, 3, 4). Calculate the zero drift according to formula (4) and take the △ with the largest absolute value as the instrument's zero drift value. A = A - Ax Calculate the range drift according to formula (5) and take the △ with the largest absolute value as the instrument's range drift value: Au = (A -A)=(An-A2 × 100% 5.4 Handling of verification results Instruments that pass the verification according to the provisions and requirements of this regulation shall be issued with a verification certificate; instruments that fail the verification shall be issued with a verification result notice and the first day of failure shall be noted. 5.5 Verification cycle The verification cycle of the instrument is 1 year. If there is any doubt about the test data of the instrument or the instrument has been replaced with a major component or repaired, it should be sent for inspection in time. Appendix A Inspection unit Instrument model Manufacturer Manufacturer Calibration environment temperature Standard gas uncertainty Appearance and functional inspection Indication error Standard gas concentration value Repeatability East value 1 Standard gas concentration value Response time Standard gas concentration value Zero point value Surplus value Zero point shift: Calibration conclusion Calibrator JJG 915—2008 Calibration record of carbon monoxide detector Measuring range Instrument No. Observation: □Normal; Others Sound and light alarm: Normal: Others Alarm concentration value: Value 1 Indication 3 Indication 2 □Unqualified Average value Indication 3 Absolute error/(μmol/mol) Indication 4 Indication 5 Response time measurement value/s Range drift: Calibration period Verifier Relative error/% Indication value 6 Repeatability Response time Appendix B Calibration items Alarm concentration value Alarm function Indication error Repeatability Response time Zero drift Range drift JJG 915—2008 Inside page of calibration certificate Technical requirements Calibration results Result judgment Appendix C Calibration items Alarm concentration value Alarm function Indication error Repeatability Ring time Zero drift Current drift Instrument failure items: J.JG915—2008 Inside page of calibration result notice Technical requirements Calibration results Result judgment5 Calibration cycle The calibration cycle of the instrument is 1 year. If there is any doubt about the test data of the instrument or the instrument has replaced major components or been repaired, it should be sent for inspection in time. Appendix A Inspection unit Instrument model Manufacturer Calibration environment temperature Uncertainty of standard gas Appearance and functional inspection Indication error Standard gas concentration value Repeatability East value 1 Standard gas concentration value Response time Standard gas concentration value Zero point value Surplus value Zero point shift: Calibration conclusion Calibrator JJG 915—2008 Calibration record of carbon monoxide detector Measuring range Instrument No. Observation: □Normal; Others Sound and light alarm: Normal: Others Alarm concentration value: Value 1 Indication 3 Indication 2 □Unqualified Average value Indication 3 Absolute error/(μmol/mol) Indication 4 Indication 5 Response time measurement value/s Range drift: Calibration period Verifier Relative error/% Indication value 6 Repeatability Response time Appendix B Calibration items Alarm concentration value Alarm function Indication error Repeatability Response time Zero drift Range drift JJG 915—2008 Inside page of calibration certificate Technical requirements Calibration results Result judgment Appendix C Calibration items Alarm concentration value Alarm function Indication error Repeatability Ring time Zero drift Current drift Instrument failure items: J.JG915—2008 Inside page of calibration result notice Technical requirements Calibration results Result judgment5 Calibration cycle The calibration cycle of the instrument is 1 year. If there is any doubt about the test data of the instrument or the instrument has replaced major components or been repaired, it should be sent for inspection in time. Appendix A Inspection unit Instrument model Manufacturer Calibration environment temperature Uncertainty of standard gas Appearance and functional inspection Indication error Standard gas concentration value Repeatability East value 1 Standard gas concentration value Response time Standard gas concentration value Zero point value Surplus value Zero point shift: Calibration conclusion Calibrator JJG 915—2008 Calibration record of carbon monoxide detector Measuring range Instrument No. Observation: □Normal; Others Sound and light alarm: Normal: Others Alarm concentration value: Value 1 Indication 3 Indication 2 □Unqualified Average value Indication 3 Absolute error/(μmol/mol) Indication 4 Indication 5Www.bzxZ.net Response time measurement value/s Range drift: Calibration period Verifier Relative error/% Indication value 6 Repeatability Response time Appendix B Calibration items Alarm concentration value Alarm function Indication error Repeatability Response time Zero drift Range drift JJG 915—2008 Inside page of calibration certificate Technical requirements Calibration results Result judgment Appendix C Calibration items Alarm concentration value Alarm function Indication error Repeatability Ring time Zero drift Current drift Instrument failure items: J.JG915—2008 Inside page of calibration result notice Technical requirements Calibration results Result judgment Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.