MT 523-1995 Explosion-proof low-energy r-r combination logging tool

Basic Information

Standard ID: MT 523-1995

Standard Name: Explosion-proof low-energy r-r combination logging tool

Chinese Name: 防爆低能r-r组合测井仪

Standard category:Coal Industry Standard (MT)

state:in force

Date of Release1996-01-15

Date of Implementation:1996-05-01

standard classification number

Standard Classification Number:Engineering Construction>>Mechanical and Electrical Manufacturing Engineering>>P82 Electronic Manufacturing Engineering

associated standards

Publication information

publishing house:China Coal Industry Press

Publication date:1996-05-01

other information

drafter:Lei Guizhong, Jia Shengmao, Guo Aihuang

Drafting unit:Xi'an Branch of China Coal Research Institute

Focal point unit:Coal Mine Special Equipment Standardization Technical Committee Coalfield Exploration Equipment Branch of the Ministry of Coal Industry

Proposing unit:Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry

Publishing department:Ministry of Coal Industry of the People's Republic of China

Introduction to standards:

This standard specifies the technical requirements, test methods, inspection rules, marking, packaging, transportation and storage of explosion-proof low-energy r-r combination logging tools. This standard is applicable to explosion-proof low-energy r-r combination logging tools for cable-free logging in horizontal and directional drilling holes underground or on the ground in coal mines. MT 523-1995 Explosion-proof low-energy r-r combination logging tools MT523-1995 Standard download decompression password: www.bzxz.net

Some standard content:

Coal Industry Standard of the People's Republic of China
MT523-1995
Explosion-proof low-energy-combination logging tool
Published on January 15, 1996
Ministry of Coal Industry of the People's Republic of China
Implementation on May 1, 1996
Coal Industry Standard of the People's Republic of China
Explosion-proof low-energy-Y combination logging tool
1 Subject content and scope of application
MT523-1995
This standard specifies the technical requirements, test methods, inspection rules, marking, packaging, transportation and storage of explosion-proof low-energy YY combination logging tool. This standard applies to explosion-proof low-energy-combination logging tools for cableless logging in horizontal and directional drilling holes underground or on the ground in coal mines. 2 Reference standards
Test method for volume resistivity and surface resistivity of solid insulating materials GB1410
GB2423.4 Basic environmental test procedures for electrical and electronic products Test Db: Alternating damp heat test method GB3836.1
General requirements for explosion-proof electrical equipment for explosive atmospheres GB3836.2 Explosion-proof electrical equipment for explosive atmospheres Flameproof electrical equipment "d" GB3836.4
Explosion-proof electrical equipment for explosive atmospheres Quality safety circuits and electrical equipment "" GB/T4942.2
Protection level of low-voltage electrical enclosures
General outline for environmental testing of electronic measuring instruments
Electronic measuring instruments
Electronic measuring instruments
3 Technical requirements
Temperature test
Transportation test
3.1 Explosion-proof low-energy-Y logging instrument (hereinafter referred to as logging instrument) shall comply with the requirements of this standard and be manufactured in accordance with the drawings and documents approved by the prescribed procedures.
3.2 The logging instrument consists of two parts: the synchronous machine and the probe. 3.3 Environmental conditions for use:
Environmental temperature: 040℃
Relative humidity: ≤95%
Environmental air pressure: (80~110)kPa
Environmental gas: gas, etc. are allowed, but corrosive gas is not allowed3.4 Electrical performance:
3.4.1 Power supply.
Power supply: composed of rechargeable battery pack:
Working voltage: ±10V, +5V, +4.5V; Working current: +10V working current is not more than 120mA, -10V working current is not more than 60mA. +5V working current is not more than 50mA, +4.5V working current is not more than 50μA. Short circuit current: +10V output short circuit current is 420mA; -10V output short circuit current is 320mA;
+5V output short circuit current is 180mA;
Approved by the Ministry of Coal Industry of the People's Republic of China on January 15, 1996, implemented on May 1, 1996
MT523-1995
+4.5V output short circuit current is 450mA;
The above values ​​are the maximum allowable values.
b. Synchronous machine
Power supply: composed of rechargeable battery pack;
Working voltage: +5V; +4.5V;
Working current: +5V working current is not more than 200mA; +4.5V working current is not more than 50μA;
Short-circuit current: +5V output short-circuit current is 430mA; +4.5V output short-circuit current is 450mA;
The above values ​​are the maximum allowable values.
Connecting cable: The length shall not exceed 0.5m. c.
3.4.2 Number of measurement channels: ≥2 channels.
3.4.3 Sampling time: ≤1s.
3.4.4 Recording word length: 8 bits for natural gamma and 16 bits for the rest. 3.4.5 Sampling method: parallel sampling.
3.4.6 Storage capacity: ≥128K.
3.4.7 Delay time: continuous or divided into different settings. Working time: The continuous normal working time of the logging instrument shall not be less than 6h. 3.4.8 Measuring range
Natural gamma: maximum count rate is 255CPS; low energy gamma: maximum count rate is 65535CPS; ground resistance: 100~20kg.
3.4.9 Ground resistance measurement accuracy: ±5%. 3.4.10 Gamma measurement sensitivity: natural gamma sensitivity: ≥600CPS/kBq; low energy gamma sensitivity: ≥400CPS/kBq. 3.5 Explosion-proof requirements:
3.5.1 The synchronous machine of the logging instrument is of intrinsic safety type, and the probe is of intrinsic safety and explosion-proof type. 3.5.2 The maximum surface temperature of the components and wires of the intrinsically safe circuit and its associated circuits and equipment under normal operation and fault conditions shall not exceed +150℃.
3.5.3 The withstand voltage and insulation resistance between the synchronous machine and the probe casing and their respective circuits after the alternating damp heat test shall comply with the relevant provisions of GB2423.4.
3.5.4 Casing.
a. The probe casing and the flameproof casing are made of seamless steel pipes with a wall thickness greater than 3mm, and should be able to withstand at least 12MPa external pressure and 0.35MPa internal pressure.
b. The synchronous machine casing is made of ABS plastic, and its bending strength should be greater than 50MPa and its tensile strength should be greater than 30MPa. No cracks are allowed in the casing after the test. The casing must be anti-static treated first, and then sprayed with anti-static paint, and the surface insulation resistance should be no greater than 1×10°0.
The protection performance of the shell should comply with the relevant provisions of the national standard GB4942.2, and the shell protection level should be IP54. c.
d. The probe should be subjected to a water pressure test, and its withstand pressure should not be less than 5MPa. 3.5.5 The panel material of the synchronous machine is anti-static hard plastic, and the surface must be treated with anti-static treatment. It should have anti-static conductive grids, and the surface insulation resistance should not be greater than 1×10°0. The panel and the shell of the synchronous machine should be firmly glued together with epoxy resin glue. 3.5.6 Power supply protection:
Each output should be connected to a current limiting resistor, and the resistance value and power consumption should comply with the relevant provisions of GB3836.4. 2
MT523-1995
3.5.7 After the power supply and circuit are debugged and qualified, including the current limiting resistor, they are all sealed with insulating materials. The glue sealing layer is required to be free of cracks and bubbles and have a certain thickness and strength.
3.5.8 Adhesives and sealant packings shall comply with the requirements of GB3836.1. 3.5.9 The withstand voltage of the isolation capacitor between the intrinsically safe and non-intrinsically safe circuits of the probe tube shall be greater than 3kV and doubled. 3.5.10 The primary and secondary of the coupling transformer that boosts the intrinsically safe power supply to 100V through the converter shall have a shielding layer and be connected to a common point. The withstand voltage between the primary and secondary and to the ground shall be greater than 1500V.
3.5.11 The cylindrical joint surface of the flameproof joint and the threaded flameproof structure shall comply with the provisions of GB3836.2. The flameproof joint and other non-flameproof joints shall have an interlocking device to ensure that they cannot be opened when powered on and cannot be powered on after opening. 3.5.12 The distance between the intrinsically safe circuit terminal and the non-intrinsically safe circuit terminal is greater than 50mm. 3.5.13 The distance between the intrinsically safe circuit and the grounding terminal is greater than 3mm. 3.5.14 The distance between the intrinsically safe circuit and the casing is greater than 3mm. 3.5.15 The insulation withstand voltage between the intrinsically safe circuit and the grounding part is greater than 500V. 3.5.16 The insulation withstand voltage between the intrinsically safe circuit and the non-intrinsically safe circuit is greater than 1500V. 3.5.17 The withstand voltage between the intrinsically safe circuit and the insulated wire used is greater than 500V. 3.6 Environmental adaptability requirements:
3.6.1 The instrument should be able to withstand the low temperature and high temperature tests under the conditions specified in Table 1, and should meet the requirements of 3.4 after the test is restored. Table 1
Test items
Low temperature operation
Low temperature storage
High temperature operation
High temperature storage
Temperature, ℃
Test time, h
Recovery time, h
3.6.2 The instrument should be able to withstand the high temperature + 40℃, 6 cycles of alternating damp heat test, and after the test, use a 500V megohmmeter to check, the insulation resistance should be not less than 0.7Ma.
3.6.3 The synchronous machine shall be able to undergo a drop test in accordance with the provisions of GB3836.1, and shall meet the requirements of 3.4 after the test. 3.6.4 After packaging, the product shall be able to withstand a transportation test of 100 km at a speed of (20-40) km/h on a third-level highway. After the test, there shall be no obvious damage on the appearance, and the electrical performance shall meet the requirements of 3.4. 4 Test method
4.1 The test conditions shall be carried out in accordance with the provisions of GB6587.1. The frequency meter of the dead position is used for detection, and the ground resistance measurement is measured with a standard precision resistor. 4.2 Performance index test Gamma count rate measurement uses 3 yuan 4.31
Gamma measurement sensitivity is tested with two standard radioactive sources of activity close to the gamma detector. 4.4 Insulation resistance detection is carried out in accordance with the provisions of GB1410. 4.5 Water pressure test The water pressure test of the probe tube is carried out by using a hydraulic press to pressurize 5MPa in the sealed tube and maintain it for 20 minutes. 4.6 The shell protection performance test shall be carried out in accordance with the provisions of GB4942.2. The sealing performance test and mechanical strength test of the introduction device shall be carried out in accordance with the provisions of GB3836.2. 4.7
The drop test shall be carried out in accordance with the provisions of GB3836.1. 4.9 The working environment temperature test shall be carried out in accordance with the provisions of GB6587.2. 4.10 The storage environment temperature test shall be carried out in accordance with the provisions of GB6587.2. The transportation test shall be carried out in accordance with the provisions of GB6587.6. 4.11
The damp heat test #
shall be carried out in accordance with the provisions of GB2423.4.
3 The flameproof performance test
MT523-1995
shall be carried out in accordance with the provisions of GB3836.2.
4.14 The intrinsic safety spark test
shall be carried out in accordance with the provisions of GB3836.4.
5 Inspection Rules
5.1 During the production and finalization process, the logging instrument must pass the inspection of this standard and obtain the explosion-proof certificate issued by the explosion-proof inspection agency designated by the Ministry of Coal Industry.
5.2 Factory Inspection:
5.2.1 Each logging instrument shall be inspected by the factory quality inspection department, and qualified products shall be issued a certificate. 5.2.2 Inspection items shall be carried out in accordance with the factory inspection requirements in Table 2. 5.2.3 All indicators of factory inspection shall meet the requirements of this standard, otherwise they will not be shipped. 5.3bzxz.net
Type test:
The logging instrument must be subjected to type test in the following cases (Table 2): When the logging instrument is finalized;
When the design, main components, materials, and processes are changed; When the explosion-proof certificate expires, it is necessary to apply to the original explosion-proof inspection agency for re-inspection; When production is resumed after one year of suspension;
Once every three years in mass production; When the quality of the logging instrument is unstable;
When the national inspection agency proposes a type test requirement. The type test should be conducted on a set of logging instruments that have passed the factory inspection. If one item except the outer shell fails in the type test, double the number of items should be drawn for re-inspection. If it still fails, the logging instrument is judged to be unqualified.
Inspection Items
Electrical Performance
Sensitivity
High Temperature Operation
High Temperature Storage
Low Temperature Operation
Low Temperature Storage
Enclosure Protection
Note: “0\ indicates the items to be inspected. Methods
Factory Inspection
Type Inspection
6 Product Marking, Packaging, Transportation and Storage
MT523-1995
6.1 Each instrument shall be fixed with Ex mark, MA mark and product nameplate in a conspicuous position. The nameplate shall include the following contents: a.
Instrument Name and Model:
Explosion-proof mark: ibdI (150℃);
Intrinsically safe rated voltage, intrinsically safe current;
Explosion-proof certificate number,
"MA" mark approval certificate;
Product number;
Manufacturing unit.
Packaging:
The logging instrument shall be packed in wooden cases before leaving the factory.
Random documents
Packing list;
Product certificate;
Product instruction manual;
Spare parts list.
6.3 Transportation:
The packaged products shall meet the requirements for water, land and air transportation. 6.4 Storage:
The logging instrument shall be stored at 0 ~40C, relative humidity not more than 85%, stored in a non-corrosive gas environment. Additional notes:
This standard was proposed by the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard is under the jurisdiction of the Coalfield Exploration Equipment Branch of the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard was drafted by the Xi'an Branch of the China Coal Research Institute. The main drafters of this standard are Lei Guizhong, Jia Shengmao, and Guo Aihuang. This standard is interpreted by the Xi'an Branch of the China Coal Research Institute. Coal
Industry Standard of the People's Republic of China
Explosion-proof Low-energy YY Combination Logging Tool
MT523-1995
Published by China Standards Press Edition
No.16, Sanlihebei Street, Fuxingmenwai, Beijing
Postal Code: 100045
Tel: 68522112
Printed by Qinhuangdao Printing Factory of China Standard Press Distributed by Beijing Distribution Office of Xinhua Bookstore Sold by Xinhua Bookstores in all regions Copyright is exclusive and no reproduction is allowed
Format 880×12301/16
Printing Sheet 1/2 Number of Words 10,000 Words
First Edition in August 1996
First Printing in August 1996
Print Quantity 1-1000
Book Number: 155066·2-10633
Heading 293-1404 Storage:
The logging instrument should be stored in an environment with a temperature of 0-40°C and a relative humidity of no more than 85% and no corrosive gas. Additional Notes:
This standard was proposed by the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard is under the jurisdiction of the Coalfield Exploration Equipment Branch of the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard was drafted by the Xi'an Branch of the China Coal Research Institute. The main drafters of this standard are Lei Guizhong, Jia Shengmao, and Guo Aihuang. This standard is interpreted by the Xi'an Branch of the China Coal Research Institute. Coal Industry Standard of the People's Republic of China
Explosion-proof Low-energy YY Combination Logging Tool
MT523-1995
Published by China Standards Press
No. 16, Sanlihebei Street, Fuxingmenwai, Beijing
Postal Code: 100045
Tel: 68522112
Printed by China Standards Press, Qinhuangdao Printing Factory, Xinhua Bookstore Distributed by Beijing Publishing House Sold by Xinhua Bookstores in all regions. Exclusive copyright. No reproduction allowed
Format 880×12301/16
Printing sheet 1/2 Number of words 10,000 words
First edition in August 1996
First printing in August 1996
Print run 1-1000
Book number: 155066·2-10633
Heading 293-1404 Storage:
The logging instrument should be stored in an environment with a temperature of 0-40°C and a relative humidity of no more than 85% and no corrosive gas. Additional Notes:
This standard was proposed by the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard is under the jurisdiction of the Coalfield Exploration Equipment Branch of the Coal Mine Special Equipment Standardization Technical Committee of the Ministry of Coal Industry. This standard was drafted by the Xi'an Branch of the China Coal Research Institute. The main drafters of this standard are Lei Guizhong, Jia Shengmao, and Guo Aihuang. This standard is interpreted by the Xi'an Branch of the China Coal Research Institute. Coal Industry Standard of the People's Republic of China
Explosion-proof Low-energy YY Combination Logging Tool
MT523-1995
Published by China Standards Press
No. 16, Sanlihebei Street, Fuxingmenwai, Beijing
Postal Code: 100045
Tel: 68522112
Printed by China Standards Press, Qinhuangdao Printing Factory, Xinhua Bookstore Distributed by Beijing Publishing House Sold by Xinhua Bookstores in all regions. Exclusive copyright. No reproduction allowed
Format 880×12301/16
Printing sheet 1/2 Number of words 10,000 words
First edition in August 1996
First printing in August 1996
Print run 1-1000
Book number: 155066·2-10633
Heading 293-140