title>GB/T 11606.8-1989 Environmental test methods for analytical instruments - Vibration test - GB/T 11606.8-1989 - Chinese standardNet - bzxz.net
Home > GB > GB/T 11606.8-1989 Environmental test methods for analytical instruments - Vibration test
GB/T 11606.8-1989 Environmental test methods for analytical instruments - Vibration test

Basic Information

Standard ID: GB/T 11606.8-1989

Standard Name: Environmental test methods for analytical instruments - Vibration test

Chinese Name: 分析仪器环境试验方法 振动试验

Standard category:National Standard (GB)

state:Abolished

Date of Release1989-03-31

Date of Implementation:1990-03-01

Date of Expiration:2007-12-01

standard classification number

Standard ICS number:Test >> 19.040 Environmental Test

Standard Classification Number:Instruments and meters>>Material composition analysis instruments and environmental monitoring instruments>>N50 Material composition analysis instruments and environmental monitoring instruments comprehensive

associated standards

alternative situation:Replaced by GB/T 11606-2007

Publication information

other information

Review date:2004-10-14

Drafting unit:Beijing Institute of Pancreatic Analytical Sciences

Focal point unit:National Technical Committee for Industrial Process Measurement and Control Standardization

Publishing department:China Machinery Industry Federation

competent authority:China Machinery Industry Federation

Introduction to standards:

GB/T 11606.8-1989 Environmental test methods for analytical instruments Vibration test GB/T11606.8-1989 Standard download decompression password: www.bzxz.net

Some standard content:

National Standard of the People's Republic of China
Environmental test method for analytical instruments
Vibration test
The method of environmental test for analytical instrumentsVibration test
1 Subject content and scope of application
GB11606.8—89
This standard specifies the grouping, test conditions and test methods for vibration tests on analytical instruments (hereinafter referred to as instruments) in non-packaged and non-working states
This standard is applicable to evaluating the adaptability of instruments to vibration and the integrity of their structures. 2 Vibration test grouping
Test stage
Resonance search
Resonance hold
Vibration cycle
Repeated resonance search
Test conditions
Frequency range Hz
Sweep rate
Amplitude peak value mm
Reported amplitude peak value mm
Hold time
Frequency range
Amplitude peak value mm
Sweep rateWww.bzxZ.net
Rate range Hz
Sweep rate
Reported amplitude peak value mm
5~33~5
1 octave per minute
5~33~5
5~33~5
Note: 1) According to the stiffness of the instrument, any value can be selected within the range of 0.075~0.19mm. 3 Test conditions
3.1.1 The instrument must be subjected to three axial vibration tests. 3.1.2 Installation of the instrument
5~55~5
0.075~0.190
1.59(5Hz≤/≤10Hz)
0.76(10Hz
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.