JB/T 9215-1999 Methods for controlling radiographic image quality

Basic Information

Standard ID: JB/T 9215-1999

Standard Name: Methods for controlling radiographic image quality

Chinese Name: 控制射线照相图像质量的方法

Standard category:Machinery Industry Standard (JB)

state:in force

Date of Release1999-06-28

Date of Implementation:2000-01-01

standard classification number

Standard ICS number:Metrology and Measurement, Physical Phenomena >> 17.240 Radiation Measurement

Standard Classification Number:Machinery>>General Machinery>>J04 Basic Standards and General Methods

associated standards

alternative situation:ZB J04002-1987

Procurement status:DIN 54109 NEQ ISO 2504 NEQ ASTM E747 NEQ

Publication information

other information

Focal point unit:National Technical Committee for Nondestructive Testing Standardization

Introduction to standards:

JB/T 9215-1999 JB/T 9215-1999 Methods for controlling radiographic image quality JB/T9215-1999 Standard download decompression password: www.bzxz.net

Some standard content:

Mechanical Industry Standard of the People's Republic of China
IR/T9215-1999
Methods for controlling the image quality of radiographic tcsting1999-06-28 Issued
State Machinery Industry Bureau
20M0-(1-01 Implementation
TBT9215-1999
This standard is not equivalent to DIN54109 Part 2 5 Metal Materials X-ray Radiography Image Quality, Image Quality Grading, and also includes [S2504 "Welding Radiographic Conditions and Background Contrast Observation Conditions - Use of Image Quality Indicator Charts", ASTM 747 "Method of Controlling Radiographic Inspection Quality with Alloy Wire Penetration Doors" and GB3323-1982 Steel Fusion Welds: Radiography and Quality Grading
This standard is a control method for 2B402-8 This standard is a revision of the method for controlling the image quality of radiographic images. During the revision, the original standard was revised: the main technical content has not changed. This standard replaces ZBJ04002-8 from the date of implementation. Appendix A of this standard is a reminder annex.
This standard was proposed by the National Technical Committee for Standardization of Nondestructive Testing. The responsible drafting unit of this standard: Shanghai Materials Research Institute: The main drafter of this standard: Chen Yuenian,
China Machinery Industry Standard
Methods for controlling the image quality of radiographic images
testing1 Scope
JB/T9215-1999bZxz.net
Total 402
This standard specifies the requirements for the quality of radiographic films when testing steel materials below 200 by radiographic method. It is a standard supporting [B9217-1997 Radiographic Operation Method]. 2 Referenced Standards
The following standards contain provisions that constitute the text of the standard by being available in the standard. When this standard is published, the versions shown are valid. All standards will be revised. Each party using this standard should discuss the feasibility of using the latest version of the following standards. JB/T790299 Image
3 Linear Image Quality Meter
This standard adopts the R'10 series in JB/T7902, and the material is marked as FF agent linear image quality meter [see Appendix A (Suggested Appendix) 4 Image Quality Meter|| tt||4.1 The mask shall be placed on the surface of the object under test on the side of the radiation source. If this is not possible, it may be placed on the surface of the object on the film side. However, it must be verified through comparative tests that the actual requirements are met. 4.2 During the inspection, the image quality meter shall be placed vertically at the end of the nearest X-ray film. 5. Number of image quality meters
5.1 Each image quality meter can only represent one valid inspection range. Within this range, the vertical change should not exceed -15%-130% of the image quality meter film. There should be at least one image quality meter image on each film. 5.2 When two films of the same or different speeds are used in the same box to expand the number of images that can be inspected in one transillumination process, several masks of corresponding specifications shall be placed on the surface of the object under test with different thicknesses according to the principle of the validity range of 5.1. 5.3 When inspecting the same type of workpieces placed in a circle or in a ring with 36° circumferential light, place one workpiece every 0°. 5.4 If the radiation source is placed at the center of a spherical penetrator and the welding line with the same geometric conditions is used for full-body light, the number of images can be obtained according to the principle of 5.3.
6 Quality level of radiography and requirements for the minimum steel diameter displayed 6.1 The quality level of radiography is divided into A level (ordinary) and B level (high sensitivity level). 6.2 The minimum wire diameter required to be displayed on the radiographic film is shown in Table 1 and Table 2. Approved by the State Machinery Industry Bureau
2000-01-01 implementation
The requirements for the penetration of steel conductors are required
The most accurate self-explanatory requirements are shown
Small diameter
Anti-requirement
The small car wire chain
Requirement
Reported wire diameter
X-ray film observation
321-copy
3B/T9215-1999
Table! Grade A requires the minimum wire diameter of the total display 6.1~8
150.1.-:70
Night 2 Grade B requires the minimum steel wire diameter of the urine tea GI-R
80 115
25 1--32
21:1-12
The blackness of A-level radiographic negatives should be equal to or greater than 1.5, and the blackness of B-level radiographic negatives should be equal to or greater than T2.0. The upper limit of the negative blackness should be within the straight line of the film sensitive light, and the background haze of the negative should be less than or equal to 0.3. The negatives should meet the requirements in the table.
Table 3 Requirements for negative viewing conditions
Low-quality test lighting is allowed
Film blackness||tt| | Mega-film light
Image display meter material square steel code,
Time record A
Appendix!
Line type quality meter with beautiful technical data
A2R[0 series image quality meter hole, line code and diameter such as A1 table Al
Line non-line number
Line number
Note the complete technical data of the line type quality meter, JBT75X2. Standard
Diameter mm
Line number
.o.125