Verification Regulation of Detector for Alcoholic Quantity from Breathing-out Gas
Some standard content:
National Metrology Verification Regulation of the People's Republic of China JJG657-2006
Detector fur Alcoholic Quantity from Breathing-out Gas2006 ~ 12 - 08 Issued
The General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China issued JJG 657—2006
Verification Regulation or Detectur forAlcoholic Quantity from Breathing -out GasJJG 657-2006
Replaces JJG657-
—1990
This regulation was approved by the General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China on December 8, 2006, and came into force on June 8, 2007.
Responsible unit: National Technical Committee for Physical and Chemical Metrology Main drafting unit: China National Institute of Metrology Ministry of Public Security Traffic Safety Product Quality Supervision and Inspection Center Participating drafting unit: Beijing Institute of Metrology and Testing Science This regulation is entrusted to the National Technical Committee for Physical and Chemical Metrology to interpret this regulation Main drafters:
Chen Yonghua
Liu Yiling
Yu Chunjun
Participating drafters:
Li Aimin
Shen Zhengsheng
JJG 657—2006
(China National Institute of Metrology)
(China National Institute of Metrology)
(Traffic Safety Product Quality Supervision and Inspection Center of the Ministry of Public Security) (Traffic Safety Product Quality Supervision and Inspection Center of the Ministry of Public Security) (Beijing Institute of Metrology and Measurement Science)
Scope·
Measuring performance requirements
Maximum indication error
3.2 Repeatability
3.3 Anti-interference ability
Memory effect
4'Functional requirements·
Blowing force
Breathing duration·||t t||Result retention or storage time
Undervoltage alarm function
5General technical requirements
Marking and accessories
Power-on inspection·
6Meter control·
Verification conditions:
Verification standard gas and equipment
Verification items·
Verification method·
Processing of test results
Test cycle
Appendix A
Appendix B
Appendix (
Appendix』
Description of ethanol standard push gas
JJG 657—2006
Verification certificate and verification result notification format Verification record format
Comparison of alcohol concentration in breath and alcohol mass concentration in blood 1
(1)
(2)
|Scope
JIG 657—206
Verification procedure for exhaled gas alcohol content detector This procedure applies to the initial verification, subsequent verification and in-use inspection of exhaled gas alcohol content detector: 2 Overview
Exhaled breath alcohol content detector (equipment store) detects the alcohol content in human exhaled gas. The detector consists of gas circuit, electronic circuit, sensor and display. The sensor types are generally electrochemical
type (fuel cell), semiconductor type and infrared type. The working principle is that ethanol passes through the sensor and is converted into current or voltage signal, and then through the analog-to-digital conversion circuit 3 Metering performance requirements
3. 1 The maximum indication value is selected when the instrument is at or below 0.2 mg/L. The maximum indication value is greater than ±0.04 mk/L. 3.2 Repeatability The indication error shall not exceed ±0.025 mg/L. In the measurement range of 0.4 mg or less, the maximum indication error shall not exceed ±0.4 mg/L. The relative error of the indication shall not exceed ±10%. 3.3 Anti-interference The relative error of the indication shall not exceed ±10%. The instrument responds to the filtered gas in the center of the air at 3%/3%. The response value should not be greater than 0.04. 3.4 Memory effect: For the test base, the response value should not be greater than 0.04. 4.1 The blowing pressure should not be greater than 3%/3. The response value of the instrument should not be greater than 0.04. 3.4 Memory effect: For the test base, the response value of the ethanol standard sampled by the instrument should not be greater than 0.04. 4.1 The blowing pressure should not be greater than 3%/3. The instrument should start sampling when the air supply pressure changes to mg/Ln. 4.2 The exhalation duration should not be greater than 240a. When the instrument is sampling, the exhaled breath is required to be interrupted and the instrument should issue an alarm signal or interrupt the measurement. 4.3 Result retention or storage time
The display or storage time of the instrument measurement result should not be less than 120sc4.4 Undervoltage alarm function
If the power supply voltage of the exhaled air flow is suddenly low during the measurement process, the instrument should alarm or indicate. 5 General technical requirements
5.1 Appearance
JJG 657--2006
The instrument should be free of damage, deformation and other phenomena that affect its normal operation. The surface of the newly manufactured instrument should be free of dull marks, obvious scratches and paint peeling. The instrument connection should be reliable and the fasteners should not be loose. 5.2 Identification and accessories
The instrument should be marked with the name, model, number, manufacturing date, manufacturing measurement instrument license identification and number of the manufacturer. The accessories should be complete and attached with the manufacturer's instruction manual and product certificate. 5.3 Power-on inspection
Each knob or button should be able to operate and control normally, and the display should be clear and correct. Measuring instrument control
The control of the instrument includes the initial calibration, subsequent calibration and in-use inspection. 6.1 Calibration conditions
6.1.1 Calibration environment conditions
Ambient temperature: (25+2); Relative humidity: not more than 80%; The calibration working environment should be free of factors that interfere with the measurement
6.2 Calibration standard gas and equipment
Ethanol standard gas: air 4 The concentration range of ethanol standard gas is: (40--500)×10-6, the expansion is not 6.2.1
The degree of certainty is not more than 3%, =2: During calibration, the unit of the standard gas should be converted into the unit of the instrument being tested. For the description of alcohol standard gas, see Appendix A.
a) Liquid organic gas gas distribution device, the expanded uncertainty of preparing ethanol standard gas is 3%, = 2: b) Alcohol breath simulator, the expanded uncertainty of preparing ethanol standard gas is 3%, -2; c) Bottled air alcohol standard gas expanded uncertainty is 3%, table = 2. 6.2.2 Anhydrous ethanol: purity should be greater than 99.7%, 6.2.3 Sampler: (0~5)L, (0-10)Lc6.2.4 Stopwatch: the division value is not greater than 0.1s
6.2.52 mg/l of the case of the burnt mixed gas, 6.2.6 clean compressed air.
South power supply: (0~24)V.
6.2.8 Le force meter: (10001100)Pa, equipped with appropriate three-way. Pressure reducing valve and anti-adsorption gas line and gas bag matching with the cylinder: 6.2.9
6.3 Inspection items
Verification items 1
Verification items
General technical requirements
Steam pressure||t t||Exhalation continuity
List of test items
First test
Continuous test
Test in use
Test items
Result retention or storage time
Undervoltage alarm function
Indication error
Repeatability
Anti-disturbance capability
Memory effect
Note: "1" indicates items to be tested,
6.4 Test method
JJG 657—2006
Table 1 (line)
First calibration
Indicates that the item may not be activated:
6.4.1 General technical requirements Perform visual and manual inspections according to 6.1·6.3. 6.4.2 Underpower alarm function
Subsequent calibration
Inspection during use
Turn on the power supply of the instrument (rated voltage value), and gradually reduce the power supply voltage. When the power supply voltage drops to the lower limit of the rated working voltage, the instrument should issue an alarm or indication. 6.4.3 Pressure
Use bottled air or nitrogen with a pressure reducing valve to slowly adjust the bed pressure. When the instrument is sampling, the pressure on the small force gauge should be greater than 240Pa.
6.4.4 Exhalation continuity
Slowly adjust the pressure. Start the sampling pump at the same time as it starts working. At 2,5. If the air flow is interrupted within 5, the instrument shall issue an alarm signal or interrupt the measurement.
6.4.5 Holding or storage time of results
Pass an ethanol standard with a mass concentration of about 0.5 into the instrument. When the instrument displays the measurement result, start the stopwatch and record the holding or storage time of the measurement result. The result shall meet the requirements of 4.3: 6.4,6 Indication error
After the instrument is powered on and preheated to stabilize, pass zero point gas and ethanol standard gas with a mass concentration of about 0.5mg/L to calibrate the zero indication of the instrument, and then pass ethanol standard gas with a mass concentration of about 0.1, 0.3, and 0.5rmg/L respectively, and record the stable indication of the instrument. The point is made three times, and the arithmetic mean of the three times is the instrument value: 0.1 rrg/L and 0.3 m/L. According to formula (1), 0.5 mg/L, the Ac element of each point is calculated by formula (2). Ac should meet the requirements of 3.1, Ac--Co
Wujin: c
Indication error;
The average value of the three indications of the instrument:
-ca×100%
The concentration of ethanol standard gas through the instrument
6.4.7 Repeatability
.JJG 657—2006
The mass concentration of ethanol standard gas is about 0.3m/L, and the instrument stable indication c is recorded. Repeat the above operation 6 times. The sum calculated by formula (3) is the repeatability (RSD) of the instrument. )?
Formula:
-th measurement value:;
-arithmetic mean of 6 measurement values
6.4.8 Anti-interference capability
General 2 m/L of air towel has been
6.4.9 Memory effect
Pass 0.3mg
standard gas into the instrument at an interval of 60:
The difference should meet the requirements of 3.4
6.5 The verification result
is marked with a verification certificate
and the unqualified items H
6.6 Verification period
Verification period light
The instrument has passed the collision
limited card
×100%
The value should meet the requirements of 3.3,
with 10.3mgl
and a verification certificate will be issued
If it falls, is repaired or replaced
Measure 0.5 Ing/L ethanol
Record the stable
value c2 and
instrument, issue a notice to the inspector, and send it for inspection in time.
METROL
Appendix A
JJG 657-2006
Description of ethanol standard gas
Because ethanol is a liquid organic matter at room temperature and normal pressure, it is easy to be adsorbed by the bottle body, valve body and gas path due to the nature of ethanol. Bottled high-pressure ethanol mixed gas (especially low concentration) has not been easy to accurately prepare and store for a long time. The measurement base range of the precision meter is generally (0-50)×. The metrological standard instrument can be: "excited gas gas preparation device", "alcohol simulator" to prepare ethanol standard gas, or bottled ethanol gas in air. The most standard is recommended according to the following requirements. The calibration cycle shall not exceed 12 years. A.1 Liquid gas preparation device should have a) carrier tape b) sampling c) preparation d) preparation e) preparation gas preparation device instruction manual and volume uncertainty 4 volume uncertainty 1 characteristic component u: 3 The linear coefficient of 0.5 should not be greater than the uncertainty of the standard gas. A.2 The alcohol device should include the instructions for use of the simulator. A.3 Bottled non-alcohol standard gas, if unpacked in China, contains self-made and promulgated air standard gas performance indicators to ensure effective traceability of the measurement value, at least including the following: maximum uncertainty of 3%, valid certificate; the prepared ethanol standard gas should also be used, and its concentration range and uncertainty approved by the metrology administrative department should meet the requirements of the regulations. MHNA
METROLA
Appendix B
Verification Items
General Technical Requirements
JJG657—2006
Verification Certificate and Verification Result Notice Inner Page FormatVerification Certificate Inner Page Format
Technical Requirements
Undervoltage Alarm Function
Blowing Pressure
Expiratory Continuity
Result Retention or Storage Time||t t||Indication error
Repeatability
Memory effect
Anti-downlift capability
Note: [Test environment; room temperature
Humidity
Atmospheric pressure
2Test basis: JIG657-2IH6 "Exhaled gas alcohol content detector" 3 Main standard instruments also used in the test:
Test results
Test items
General technical requirements||tt| |Undervoltage alarm function
Blowing pressure
Exhalation continuity
Result retention or storage time
Indication error
Repeatability
Memory effect
Anti-interference ability
Indicate unqualified items:
Note: 1. Verification environment: room temperature
JJG657-2006
Verification result notification page format
Technical requirements Request
and condensation
Atmospheric island
2 Calibration basis: J657—2006 Gas Alcohol Content Detector》 3 Main standard instruments used in calibration:
Calibration results
Appendix C
Certification date:
Instrument name:,
Inspection unit:
Calibration basis:
Calibration standards and main supporting equipment
JIG 657—2006
Calibration record format
Original record number
Daily temperature
Certificate number
Relative humidity
Factory number
SHOHNS
Manufacturer:
, General technical requirements:
. , indication error, weight
standard value
150/10A
to, anti-down excellent ability:
4. Memory effect:
5. Blow (pressure:
6. Exhalation continuity:
Hand disturbance gas roller
t. Result retention or storage time
8. Undervoltage alarm function:
Verification conclusion:
Instrument indication
METROLO
Verifier
Verifier3mg
standard gas, interval 60:
the difference should meet the requirements of 3.4
6.5
of the calibration result
and indicate the unqualified items
6.6 Calibration period
Calibration period light
The instrument passes through the collision
limit|
×100%
value, it should meet the requirements of 3.3,
ethanol standard gas, record the indicated value c,, 10.3mgl
issue a calibration certificate
, repair or replace
measure 0.5 Ing/L ethanol
record the stable value c2 and
instrument, issue a notice of failure to the inspector, and send it for inspection in time.
METROL
Appendix A
JJG 657-2006
Description of ethanol standard gas
Because ethanol is a liquid organic substance at normal temperature and pressure, it is easily adsorbed by the bottle, valve body and gas path due to the properties of ethanol. Bottled high-pressure ethanol mixed gas (especially low concentration) has not been easy to accurately prepare and store for a long time. The measurement base range of the precision meter is generally (0-50)×. The metrological standard instrument can be: "excited gas gas preparation device", "alcohol simulator" to prepare ethanol standard gas, or bottled ethanol gas in air. The most standard is recommended according to the following requirements. The calibration cycle shall not exceed 12 years. A.1 Liquid gas preparation device should have a) carrier tape b) sampling c) preparation d) preparation e) preparation gas preparation device instruction manual and volume uncertainty 4 volume uncertainty 1 characteristic component u: 3 The linear coefficient of 0.5 should not be greater than the uncertainty of the standard gas. A.2 The alcohol device should include the simulator instructions for use. A.3 Bottled non-alcohol standard gas, if unpacked in China, contains self-made and promulgated air standard gas performance indicators to ensure effective traceability of the measurement value, at least including the following: maximum uncertainty of 3%, valid certificate; the prepared ethanol standard gas should also be used, and its concentration range and uncertainty approved by the metrology administrative department should meet the requirements of the regulations. MHNA
METROLA
Appendix B
Verification Items
General Technical Requirements
JJG657—2006
Verification Certificate and Verification Result Notice Inner Page FormatVerification Certificate Inner Page Format
Technical Requirements
Undervoltage Alarm Function
Blowing Pressure
Expiratory Continuity
Result Retention or Storage Time||t t||Indication error
Repeatability
Memory effect
Anti-downlift capability
Note: [Test environment; room temperature
Humidity
Atmospheric pressure
2Test basis: JIG657-2IH6 "Exhaled gas alcohol content detector" 3 Main standard instruments also used in the test:
Test results
Test items
General technical requirements||tt| |Undervoltage alarm function
Blowing pressure
Exhalation continuity
Result retention or storage time
Indication error
Repeatability
Memory effect
Anti-interference ability
Indicate unqualified items:
Note: 1. Verification environment: room temperature
JJG657-2006
Verification result notification page format
Technical requirements Request
and condensation
Atmospheric island
2 Calibration basis: J657—2006 Gas Alcohol Content Detector》 3 Main standard instruments used in calibration:
Calibration results
Appendix C
Certification date:
Instrument name:,
Inspection unit:
Calibration basis:
Calibration standards and main supporting equipment
JIG 657—2006
Calibration record format
Original record number
Daily temperature
Certificate number
Relative humidity
Factory number
SHOHNS
Manufacturer:
, General technical requirements:
. , indication error, weight
standard value
150/10A
to, anti-down excellent ability:
4. Memory effect:
5. Blow (pressure:
6. Exhalation continuity:
Hand disturbance gas roller
t. Result retention or storage time
8. Undervoltage alarm function:Www.bzxZ.net
Verification conclusion:
Instrument indication
METROLO
Verifier
Verifier3mg
standard gas, interval 60:
the difference should meet the requirements of 3.4
6.5
of the calibration result
and indicate the unqualified items
6.6 Calibration period
Calibration period light
The instrument passes through the collision
limit|
×100%
value, it should meet the requirements of 3.3,
ethanol standard gas, record the indicated value c,, 10.3mgl
issue a calibration certificate
, repair or replace
measure 0.5 Ing/L ethanol
record the stable value c2 and
instrument, issue a notice of failure to the inspector, and send it for inspection in time.
METROL
Appendix A
JJG 657-2006
Description of ethanol standard gas
Because ethanol is a liquid organic substance at normal temperature and pressure, it is easily adsorbed by the bottle, valve body and gas path due to the properties of ethanol. Bottled high-pressure ethanol mixed gas (especially low concentration) has not been easy to accurately prepare and store for a long time. The measurement base range of the precision meter is generally (0-50)×. The metrological standard instrument can be: "excited gas gas preparation device", "alcohol simulator" to prepare ethanol standard gas, or bottled ethanol gas in air. The most standard is recommended according to the following requirements. The calibration cycle shall not exceed 12 years. A.1 Liquid gas preparation device should have a) carrier tape b) sampling c) preparation d) preparation e) preparation gas preparation device instruction manual and volume uncertainty 4 volume uncertainty 1 characteristic component u: 3 The linear coefficient of 0.5 should not be greater than the uncertainty of the standard gas. A.2 The alcohol device should include the simulator instructions for use. A.3 Bottled non-alcohol standard gas, if unpacked in China, contains self-made and promulgated air standard gas performance indicators to ensure effective traceability of the measurement value, at least including the following: maximum uncertainty of 3%, valid certificate; the prepared ethanol standard gas should also be used, and its concentration range and uncertainty approved by the metrology administrative department should meet the requirements of the regulations. MHNA
METROLA
Appendix B
Verification Items
General Technical Requirements
JJG657—2006
Verification Certificate and Verification Result Notice Inner Page FormatVerification Certificate Inner Page Format
Technical Requirements
Undervoltage Alarm Function
Blowing Pressure
Expiratory Continuity
Result Retention or Storage Time||t t||Indication error
Repeatability
Memory effect
Anti-downlift capability
Note: [Test environment; room temperature
Humidity
Atmospheric pressure
2Test basis: JIG657-2IH6 "Exhaled gas alcohol content detector" 3 Main standard instruments also used in the test:
Test results
Test items
General technical requirements||tt| |Undervoltage alarm function
Blowing pressure
Exhalation continuity
Result retention or storage time
Indication error
Repeatability
Memory effect
Anti-interference ability
Indicate unqualified items:
Note: 1. Verification environment: room temperature
JJG657-2006
Verification result notification page format
Technical requirements Request
and condensation
Atmospheric island
2 Calibration basis: J657—2006 Gas Alcohol Content Detector》 3 Main standard instruments used in calibration:
Calibration results
Appendix C
Certification date:
Instrument name:,
Inspection unit:
Calibration basis:
Calibration standards and main supporting equipment
JIG 657—2006
Calibration record format
Original record number
Daily temperature
Certificate number
Relative humidity
Factory number
SHOHNS
Manufacturer:
, General technical requirements:
. , indication error, weight
standard value
150/10A
to, anti-down excellent ability:
4. Memory effect:
5. Blow (pressure:
6. Exhalation continuity:
Hand disturbance gas roller
t. Result retention or storage time
8. Undervoltage alarm function:
Verification conclusion:
Instrument indication
METROLO
Verifier
Verifier3. Bottled empty 4. Non-alcohol standard gas, if unpacked in China, the self-made and promulgated air standard gas quality performance must be guaranteed to have effective traceability of the value, and should at least include: maximum uncertainty (>3%), valid certificate; the prepared ethanol standard gas should also be used, and its concentration range and uncertainty approved by the metrology administrative department must meet the requirements of the regulations. MHNA
METROLA
Appendix B
Verification Items
General Technical Requirements
JJG657—2006
Verification Certificate and Verification Result Notice Inner Page FormatVerification Certificate Inner Page Format
Technical Requirements
Undervoltage Alarm Function
Blowing Pressure
Expiratory Continuity
Result Retention or Storage Time||t t||Indication error
Repeatability
Memory effect
Anti-downlift capability
Note: [Test environment; room temperature
Humidity
Atmospheric pressure
2Test basis: JIG657-2IH6 "Exhaled gas alcohol content detector" 3 Main standard instruments also used in the test:
Test results
Test items
General technical requirements||tt| |Undervoltage alarm function
Blowing pressure
Exhalation continuity
Result retention or storage time
Indication error
Repeatability
Memory effect
Anti-interference ability
Indicate unqualified items:
Note: 1. Verification environment: room temperature
JJG657-2006
Verification result notification page format
Technical requirements Request
and condensation
Atmospheric island
2 Calibration basis: J657—2006 Gas Alcohol Content Detector》 3 Main standard instruments used in calibration:
Calibration results
Appendix C
Certification date:
Instrument name:,
Inspection unit:
Calibration basis:
Calibration standards and main supporting equipment
JIG 657—2006
Calibration record format
Original record number
Daily temperature
Certificate number
Relative humidity
Factory number
SHOHNS
Manufacturer:
, General technical requirements:
. , indication error, weight
standard value
150/10A
to, anti-down excellent ability:
4. Memory effect:
5. Blow (pressure:
6. Exhalation continuity:
Hand disturbance gas roller
t. Result retention or storage time
8. Undervoltage alarm function:
Verification conclusion:
Instrument indication
METROLO
Verifier
Verifier3. Bottled empty 4. Non-alcohol standard gas, if unpacked in China, the self-made and promulgated air standard gas quality performance must be guaranteed to have effective traceability of the value, and should at least include: maximum uncertainty (>3%), valid certificate; the prepared ethanol standard gas should also be used, and its concentration range and uncertainty approved by the metrology administrative department must meet the requirements of the regulations. MHNA
METROLA
Appendix B
Verification Items
General Technical Requirements
JJG657—2006
Verification Certificate and Verification Result Notice Inner Page FormatVerification Certificate Inner Page Format
Technical Requirements
Undervoltage Alarm Function
Blowing Pressure
Expiratory Continuity
Result Retention or Storage Time||t t||Indication error
Repeatability
Memory effect
Anti-downlift capability
Note: [Test environment; room temperature
Humidity
Atmospheric pressure
2Test basis: JIG657-2IH6 "Exhaled gas alcohol content detector" 3 Main standard instruments also used in the test:
Test results
Test items
General technical requirements||tt| |Undervoltage alarm function
Blowing pressure
Exhalation continuity
Result retention or storage time
Indication error
Repeatability
Memory effect
Anti-interference ability
Indicate unqualified items:
Note: 1. Verification environment: room temperature
JJG657-2006
Verification result notification page format
Technical requirements Request
and condensation
Atmospheric island
2 Calibration basis: J657—2006 Gas Alcohol Content Detector》 3 Main standard instruments used in calibration:
Calibration results
Appendix C
Certification date:
Instrument name:,
Inspection unit:
Calibration basis:
Calibration standards and main supporting equipment
JIG 657—2006
Calibration record format
Original record number
Daily temperature
Certificate number
Relative humidity
Factory number
SHOHNS
Manufacturer:
, General technical requirements:
. , indication error, weight
standard value
150/10A
to, anti-down excellent ability:
4. Memory effect:
5. Blow (pressure:
6. Exhalation continuity:
Hand disturbance gas roller
t. Result retention or storage time
8. Undervoltage alarm function:
Verification conclusion:
Instrument indication
METROLO
Verifier
Verifier