Some standard content:
National Metrology Technical Specification of the People's Republic of China JJF1110—2003
Calibration Specificatiun tur Construction Quality Tesler Sets2003-09-23Promulgated
Implementation on 2003-12-23
Promulgated by the General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China JJK1110—2003
Calibration Specification for Construction Quality Test Sets
Calibration Specification for Construction Quality Test SetsJJF 1110--2003
This standard was approved by the General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China on September 23, 2003 and came into effect on December 23, 2003.
Responsible unit:
Main drafting unit:
Participating drafting unit:
Enterprise National Length Metrology Technical Committee Shenzhen Metrology Quality Inspection Institute
Shanxi Metrology Supervision and Verification Testing Institute
Wenzhou Nanfang Building Instrument Factory
This specification is entrusted to the responsible unit for interpretation
Main drafters of this specification:www.bzxz.net
Yu Jiping
Shi Xuexiang
Participating drafter!
Yang Shicong
JJF 1110 --2003
{Shenzhen Metrology and Quality Inspection Institute
(Shanxi Metrology Supervision and Verification Test Institute) (Shenzhen Order Quality Inspection Institute) (Wenzhou Nanfang Building Instruments)
References
4 Metrology characteristics
Verticality detection
4.2 Internal and external right angle detection ruler
Diagonal detection ruler
Feeler gauge
Hundred grid
4.6 Slope ruler
5 Calibration equipment
Environmental conditions
5.2 Standard instruments and other equipment||t t||6 Calibration items and calibration methods
Vertical test piece
Inside and outside right angle test ruler
6.3 Scale length deviation of diagonal test ruler
.4 Thickness deviation of feeler gauge
JJF1111·2003
6.5 Scale for deviation of length and width of the measuring rod
7 Expression of calibration results
8 Time interval of calibration
Appendix A Format of calibration results
Appendix B Simplified diagram of calibration device
1 Scope
JJF 1110—2003
Calibration specification for construction engineering quality detector group This specification is applicable to the calibration of verticality detection ruler, internal and external right angle detection ruler, opposite detection ruler, shape ruler, grid and slope ruler used for construction engineering quality inspection. The types of various instruments are shown in Figure 1. Schematic diagram of construction engineering quality detector group
: 1. Verticality detection ruler, 2. Energy measurement: 3. Level reference: 4. Fixed fulcrum: 5. New fulcrum: 6. -Inside and outside = light measurement scale: -Position cloud table; B. Folding scale; -A pair of single-path keys: 10-Stretching scale and element scale: I1-Error scale: 12-Grid: 17-Slope scale: 11-Bracket table: 15-Water parts included 16-Adjustment sign vehicle 4,: 7-Front music fat: 18-Back cell 2 References
The provisions contained in the following documents constitute the provisions of this specification through reference in this specification. When this specification is published, the versions shown are valid. All standards, specifications or regulations will be revised. All parties using this specification will discuss the possibility of using the latest version of the following documents. JJF1001—1998
CB/T15481—2000
3Overview
Terms and definitions
General requirements for inspection and laboratory capabilitiesThe construction engineering quality inspection system consists of a verticality inspection ruler, an internal and external vertical increase inspection ruler, an alignment inspection ruler, a shape-changing feeler gauge, a grid scale and a slope gauge. It is also commonly used to measure the verticality and flatness of building walls, the internal and external vertical increase, the alignment of door frames, the length difference between lines, gaps, the saturation of brick mortar, etc.4 Measuring characteristics
4. Upper verticality detection ruler
The straightness of the working surface shall not exceed 0.5mm over a length of 2m4.1.11
The vertical distance difference between the end faces of the three fulcrums and the working surface shall not exceed 0.1u. The indication value variation shall not exceed 0.2mm over a length of 2m4.1.4
Indication error shall be 0.5mm/2m
The straightness of the working surface of the internal and external right angle detection ruler
shall not exceed 0.1m.
The parallelism of each two working surfaces is not greater than 0.15mm, 4.2.33
Zero value error: ±(,1 graduation value
4.2.4Minimum value error: ±0.5mm/50mm
4.3Diagonal detection ruler
The ruler length difference is ±0.5mml
4.4Flange feeler gauge
The degree deviation is ±0.2m.
4.5Hundred grid
The relative deviation of length and width dimensions is =0.5%4.6Slope ruler
4.6.1The zero value error of the level is not greater than 0.2 graduation value. 4.6.2:Value error ±0.5mmrm
Note: The change is not determined by the change, and the measurement is advanced Characteristics index This is for reference 5 Calibration conditions
5.1 Environmental conditions
Overflow: (20±10):
Humidity: not less than 80%RH.
5.2 Standards and other equipment
Flat plate or flat ruler; table ruler, verticality detection ruler indication error calibration device; caliper: internal and external right angle detection ruler indication error calibration device: height gauge: steel ruler: 5 equal mother blocks, 6 Calibration items and calibration methods
Before calibration, it should be confirmed that there is no shadow combing, the correct implementation of the calibration and the appearance defects and mechanical failures of the calibration results, 6.1 Verticality detection ruler
6..11. Straight line film of the work page
Put the working surface of the vertical run detection ruler flat on a flat plate or flat ruler, and select a 0.5mm thick Gu The ruler is used to test the quality of the flat plate or flat ruler and the work surface. When it cannot be inserted, its straightness reaches the index of 4,1.1, 2
JJFJEO20N
6.12 The temperature difference of the three supporting points is measured with a cloud ruler until the main vertical distance of the three supporting points of the ruler is measured to the main point. The maximum and minimum differences are as shown in the manufacturer's specification.
6.1.3 Also as shown in the variable or
Egg Figure 2, install this straightness detection ruler on the calibration device of the general straightness detection ruler (hereinafter referred to as the calibration device 1), and measure the micrometer head of the calibration device to the zero position: then read the pointer of the indicator table of the verticality detection ruler to the zero position. Rotate the micrometer head forward (backward) to make the indicator table deviate from the zero position, and move the micrometer head to make the micrometer head refocus on zero and read the pointer on the indicator table. 6.1.4 Small error
Install the calibrated calibrated scale on the calibration device, turn the calibration device’s micrometer head, adjust the pointer to the zero position, and set the micrometer reading to zero.
Use 5 accurate points, 3, 9, 12, 15: Before turning the micrometer head of the calibration device, the calibration step should be to check the zero position of the scale indication: it should be turned micrometer, if the scale is zeroed, it should be more wrong:
Calibration, turn the micrometer head in turn, and get the standard value relative to the calibration point: the reading of the calibrated scale is equal to the standard value, and the reading minus the standard value is the actual indication error.
6.2 External inspection
6.2.1 The vertical load of the working mountain
. Release the epidemic type! :
5-positive stimulation,
4 sensitive control scale" steam surface
Put the external inspection scale on the working mountain and put it on the flat plate L, the pressure is 1nm, the test frame length and the working mountain are multiplied by 1nm, and the linear measurement reaches the index of 21. .2.3 The vertical measurement of the two working mountains
The calipers are used to measure the distance between the work seat and the work seat respectively: measure to one point, and the difference between the minimum values of the calipers is the vertical measurement of the working mountain.
6.2.3 Zero value error
Put the calibrated inside and outside right angle detection ruler and the inside and outside right angle measurement ruler calibration device (hereinafter referred to as calibration device) on the flat surface as shown in Figure 3, ensure that the calibrated ruler fits tightly with the flat surface and the calibration device Ⅱ, and then observe the indicator of the calibrated ruler. At this time, the indicator indicates the zero value error of the outside angle: then put the calibrated ruler as shown in Figure 4, ensure that the calibrated ruler and the calibration device Ⅱ fit tightly, and at this time, the indicator indicates the zero value error of the inside angle. Take the maximum value as the zero error of the calibrated inside and outside right angle detection ruler,
6.2.4 Indication error
If the ruler is as shown in Figure 3, the calibration needs to be supplemented with the indication error of the scale: the ruler is as shown in Figure 4. Note the error of the calibration value on the right side: When calibrating, adjust the measuring head of the calibration device to zero, select 0, 11, 3, 5mm, 7mm as 3
IJF 1110--2603
calibration points. Rotate the micrometer head according to the model, give the standard value relative to the calibration point, read the digital value from the indicator of the calibrated ruler, the reading minus the standard value is the indication error of the corresponding point, so 3
1-plate, 2-adjustable ruler
3-calibration value II: 4-position head
. 3 Diagonal measuring ruler scale length deviation 4
[deck: 2-precision ruler:
light measurement-probe
Use steel ruler to directly measure the length of each scale of the measuring ruler: the measurement interval is 20mm. The reading of the steel ruler minus the nominal value of the ruler is the ruler length deviation. 6.4 Thickness deviation of the flat ruler
Take the upper ruler as the standard, and measure at intervals of 3mm. When measuring, place the height caliper and the feeler gauge on the flat plate, adjust the height caliper up and down, and put its claws to touch the corresponding scale line of the flat ruler. Read the highest value from the height caliper. The reading value minus the nominal value is the thickness deviation of the flat ruler. 6.5 Relative deviation of the length and width of the white grid Use a steel ruler to measure the length and width of the white grid respectively, subtract the nominal value from the reading value, and then divide it by the nominal value to get the relative deviation of the length and width of the white grid. 6.6 Slope ruler
6.1 Zero value error of level
Put the slope ruler flat on a flat plate, adjust the adjusting screw to make the - end of the water bubble align with the middle scale line on one side; then rotate the slope ruler horizontally 180°: put it in the same position, and check whether the position of the level bubble on the side deviates from the middle scale line. Half of the deviation is the zero value error of the level. 6.6.2 Error correction
Use the block as the standard, and calibrate it by 5mm
First Level the flat plate so that the angle between the plate and the water surface is less than 0.mmm. Then place the slope gauge on flat plate 1 and use the adjusting screw to adjust the bubble of the spirit level to the middle position. The pointer of the slope gauge indicator should point to zero. After adjustment, place a 5-diameter plate under the front plane and use the adjusting screw to adjust the bubble of the spirit level of the slope gauge to the middle position. Note the reading of the indicator. The reading minus the standard value is the indication error of the slope gauge at that point. Replace other gauge blocks in turn and measure the indication difference at other points. The certificate shall include at least the following information: 8) Title: Calibration Certificate: b) Name and address of the laboratory: J.1H1110-2003; unique identification of the calibration certificate (such as serial number) and identification of each page to ensure that the page can be identified as part of the calibration certificate, as well as a clear identification of the end of the calibration certificate; d) Name and address of the manufacturer; ) Understand the use of this specification as the basis for calibration: 1) Description, status and clear identification of the calibration instrument; g) Calibration period; b) Calibration results: Signatures of the calibrator, recalibrator and approver; ) Environmental conditions; k) Evidence of the source of the measurement inaccuracy (or) the calibration results comply with this specification or the clauses: 1) Evidence of the source of the measurement inaccuracy. The recommended format of the calibration results is shown in Appendix A: Recalibration time interval. The recommended time interval should not exceed two months. Appendix A
1.Verticality measuring ruler
Linearity of working surface:
JJF1118—2003
Calibration result format
Calibration result
Vertical distance between the three-dimensional surface and the working surface: deviation:
Difference error:
Difference error Uncertainty of calibration result:
2.Internal and external right angle measuring ruler
Straightness of working surface:
Parallelism of each two working surfaces:||tt| |Scientific error:
Indication error:
Uncertainty of calibration result: 3. Diagonal detection
Scale length deviation:
4. Model scale
Thickness deviation:
5. Grid
Relative deviation of length and width dimensions
6. Zero value deviation of measured scale
Level group:
Indication error:
Indication errorUncertainty of calibration result:
Appendix B
Calibration device I See Figure H.1,
1. Material: rock;
2 with 50mm1 drum measuring head
JJF1110-2003
Simplified diagram of the calibration device
1 3080
Calibration device II: See B.2.
Instructions:
1. C plane consideration: 0.005;
2.7-perpendicular point of the work surface: 0.U1;
3 Resistance: 3;
4, with 250m total frequency head
JJF1110-20031H1110—2003
Unique identification of the calibration certificate (such as serial number) and identification of each page to ensure that the page can be identified as part of the calibration certificate, as well as a clear indication of the end of the calibration certificate; d) Name and address of the manufacturer;
) Principles of the use of this specification: 1) Description, condition and clear identification of the calibration equipment; g) Calibration period;
b) Calibration results:
Signatures of the calibrator, calibrator and approver; ) Environmental conditions;
k) Evidence of the source of the measurement inaccuracy and (or) the calibration results comply with this specification or the clauses: 1) Evidence of the source of the measurement inaccuracy. The recommended format of the calibration results is shown in Appendix A: Recalibration time interval
The recommended time interval should not exceed two months. Appendix A
1.Verticality measuring ruler
Linearity of working surface:
JJF1118—2003
Calibration result format
Calibration result
Vertical distance between the three-dimensional surface and the working surface: deviation:
Difference error:
Difference error Uncertainty of calibration result:
2.Internal and external right angle measuring ruler
Straightness of working surface:
Parallelism of each two working surfaces:||tt| |Scientific error:
Indication error:
Uncertainty of calibration result: 3. Diagonal detection
Scale length deviation:
4. Model scale
Thickness deviation:
5. Grid
Relative deviation of length and width dimensions
6. Zero value deviation of measured scale
Level group:
Indication error:
Indication errorUncertainty of calibration result:
Appendix B
Calibration device I See Figure H.1,
1. Material: rock;
2 with 50mm1 drum measuring head
JJF1110-2003
Simplified diagram of the calibration device
1 3080
Calibration device II: See B.2.
Instructions:
1. C plane consideration: 0.005;
2.7-perpendicular point of the work surface: 0.U1;
3 Resistance: 3;
4, with 250m total frequency head
JJF1110-20031H1110—2003
Unique identification of the calibration certificate (such as serial number) and identification of each page to ensure that the page can be identified as part of the calibration certificate, as well as a clear indication of the end of the calibration certificate; d) Name and address of the manufacturer;
) Principles of the use of this specification: 1) Description, condition and clear identification of the calibration equipment; g) Calibration period;
b) Calibration results:
Signatures of the calibrator, calibrator and approver; ) Environmental conditions;
k) Evidence of the source of the measurement inaccuracy and (or) the calibration results comply with this specification or the clauses: 1) Evidence of the source of the measurement inaccuracy. The recommended format of the calibration results is shown in Appendix A: Recalibration time interval
The recommended time interval should not exceed two months. Appendix A
1.Verticality measuring ruler
Linearity of working surface:
JJF1118—2003
Calibration result format
Calibration result
Vertical distance between the three-dimensional surface and the working surface: deviation:
Difference error:
Difference error Uncertainty of calibration result:
2.Internal and external right angle measuring ruler
Straightness of working surface:
Parallelism of each two working surfaces:||tt| |Scientific error:
Indication error:
Uncertainty of calibration result: 3. Diagonal detection
Scale length deviation:
4. Model scale
Thickness deviation:
5. Grid
Relative deviation of length and width dimensions
6. Zero value deviation of measured scale
Level group:
Indication error:
Indication errorUncertainty of calibration result:
Appendix B
Calibration device I See Figure H.1,
1. Material: rock;
2 with 50mm1 drum measuring head
JJF1110-2003
Simplified diagram of the calibration device
1 3080
Calibration device II: See B.2.
Instructions:
1. C plane consideration: 0.005;
2.7-perpendicular point of the work surface: 0.U1;
3 Resistance: 3;
4, with 250m total frequency head
JJF1110-2003
Tip: This standard content only shows part of the intercepted content of the complete standard. If you need the complete standard, please go to the top to download the complete standard document for free.